Contact an Account ManagerSimultaneous measurement of endogenous steroid hormones and their metabolites with LC-MS/MS in faeces of a New World primate species, Cebus capucinus
Summary
Type: SPE
Application #: 305710
Detection:
Conditions: Column conditioning: 1 mL methanol, then 1 mL water and at last 1 mL phosphate buffer (pH 6.9); (Do not let run the column dry!)
Sample aspiration: 500 µL pretreated sample was passed through the column by vacuum
Washing: 3 mL water followed by 1 mL 40 % methanol
Elution: 1 mL methanol, then 1 mL ethyl acetate collected in a single fraction
Authors: Weltring, A. et al.
Source: Physiology & Behavior 105 (2012) 510–521
Details
Legend: